Select your language

Ammonium- und Nitratsensor: ISE-Kombisensor - WTW

PDFPrintEmail
Ammonium and nitrate sensor: ISE combination sensor
Product Weight 0,200   Kg
Product Length 20,0   cm
Product Width 5,0   cm
Product Height 5,0   cm

For more detail please ask.
 
 
Produktbeschreibung:
Description

ISE combination sensor for ammonium and nitrate - WTW

Digital ISE sensor for NH4 and NO3

  • ISE: Ion selective electrode
  • Digital ISE sensor for NH4 and NO3
    Reagent-free ammonium and nitrate ion-selective measurement with automatic compensation of potassium/chloride with the VARiON® Plus 700 IQ:
    • Ammonium and nitrate measurement directly in the process
    • Simultaneous compensation of potassium, chloride
    • No sampling, reagent-free
    • Calibration-free , long-term stable
The particularly stable electrodes guarantee a long service life and easy cleaning. To measure, simply install the corresponding electrode into the IQ SENSOR NET sensor; everything else happens automatically. The values ​​that have already been compensated are shown in the measured value display, either with potassium or chloride output. These values ​​are also directly available for signal output via 0/4–20 mA output or digital output (Profibus, Modbus RTU, Ethernet/IP, Modbus TCP).
 
   Ammonium measurement  Nitrate measurement

Measuring principle

Electrochemical

Maximum configuration of a sensor

Common reference electrode, two measuring electrodes, one compensation electrode

Integrable measuring chains:
reference


VARiON ® Plus  Ref

measuring electrode

VARiON® Plus  NH _

VARiON ® Plus  NO 3

compensation

VARiON ® Plus  K

VARiON ® Plus  Cl

Measuring ranges/resolution


NH 4 -N :
1 ... 2,000 mg/l / 1 mg/l;
0.1 ... 100 mg/l / 0.1 mg/l

NH + :
1 ... 2,580 mg/l / 1 mg/l;
0.1 ... 129.0 mg/l / 0.1 mg/l

NO 3 -N:
1 … 1,000 mg/l / 1 mg/l;
0.1 ... 100 mg/l / 0.1 mg/l

NO 3 -:
5 … 4500 mg/l / 1 mg/l;
0.5 ... 450.0 mg/l / 0.1 mg/l

Compensation areas

+ :
0.1 ... 1,000 mg/l / 0.1 mg/l

Cl - :
0.1 ... 1,000 mg/l / 0.1 mg/l

Measurement accuracy in laboratory standards

± 5% of measured value ± 0.2 mg/l in standard solutions

Calibration types

Matrix adjustment against any reference value, 2-point calibration with combined standards possible

Downtime (typical)

Reference electrode: 18 months, measuring and compensation electrode: 18 months for typical application (municipal sewage treatment plants)

Temperature measurement
and compensation

Integrated NTC,
range 0 °C … +40 °C,
accuracy ±0.5 K, resolution 0.1 K, t 95  < 20 s

Compressive strength

Maximum 0.2 bar (including sensor connection cable, with built-in measuring chains)

Environmental conditions

Operating temperature: 0 °C … +40 °C, storage temperature: 0 °C … +40 °C

Electrical connection

2-core shielded cable with quick release on the sensor

Electromagnetic compatibility

EN 61326, Emission: Class B, FCC Class A
Immunity for essential operation

Test mark

CE

mechanics

Housing shaft: V4A stainless steel 1.4571
Protective basket: POM,
temperature sensor: V4A stainless steel 1.4571
Protection class: IP 68 (0.2 bar, with built-in measuring chains)
Measuring chain holder: POM

Weight

Approx. 670 g (without measuring chains, without sensor connection cable)

guarantee

VARiON ® Plus  700 IQ: 2 years
Electrodes: 1 year for material defects in accordance with § 10 General Terms and Conditions

data sheet  

    WTW 107060 VARiON Plus 700 IQ ISE sensor

    Units in box: 1

     
    © Borgmann-Aquaponik-Hydroponik.ch